The following is LENPURE tells you about the maintenance of the refrigeration system of the high and low temperature test chamber, please visit: Because the price of high and low temperature test chamber is basically composed of electrical, refrigeration and mechanical systems. Therefore, once there is a problem with the equipment, it is necessary to comprehensively check and comprehensively analyze the entire equipment system. Generally speaking, the process of analysis and judgment can be carried out first, after the external factors are eliminated first, and then the system is decomposed based on the fundamental fault phenomenon. After the comprehensive analysis and judgment of the system, or you can also use the backward method to find the cause of the failure. First, check whether there is a problem with the electrical system according to the electrical wiring diagram, and finally find out whether it is a problem with the refrigeration system. Before you can understand the cause of the failure, you must not blindly disassemble or replace parts to avoid unnecessary trouble. Our company generally recommends placing the high and low temperature test chamber at a temperature of 8 ° C ~ 28 ° C. For laboratories that do not have this condition, an appropriate air conditioner must be equipped. Adhere to the professional management and maintenance of special personnel. Qualified units should send special personnel to the supplier's factory for training and learning from time to time to obtain professional maintenance and repair experience and capabilities. The condenser is cleaned once every three months. If the compressor is cooled by air, the condenser fan should be regularly repaired and the condenser should be decontaminated and dedusted to ensure good ventilation and heat exchange performance. For compressors using water cooling, in addition to ensuring the inlet water pressure and inlet water temperature, the corresponding flow rate must also be guaranteed, and the inside of the condenser should be cleaned and descaled regularly to obtain its continuous heat exchange performance. The evaporator should also be cleaned regularly in the test chamber, because the cleanliness of the samples varies, and a lot of dust and small particles will condense on the evaporator under the action of forced air circulation, so it should be cleaned regularly. There are also waterway and humidifier cleaning. If the waterway is not smooth and the humidifier is fouled, it will cause the humidifier to dry and burn, which may damage the humidifier. Anti-Sensitivity Toothpaste,Anti-Sensitive Dental Paste,Strengthen Teeth Whitening Gel,Anti Bacterial Toothpaste Huai 'an Bst Trade Co., LTD , https://www.bstoral.com